VICTOR 6310B DC Low Resistance Tester Micro-ohmmeter Ohmmeter LCD Display
Notice:
The standard package don't include batteries.Because,the international air shipping forbid to ship liquid,powder and batteries,we can not put them in the package box.Hope you can understand buy them in your local store or supermarket.
At the same time, please forgive us that we are unable to provide the English user manual. If you need it, please contact the customer service after purchase.
Thanks for your support!
DC low resistance meter, also known as micro-ohmmeter, ohmmeter, DC resistance tester, adopts microprocessor technology, four-wire method for testing, safe, precise and reliable. Mainly used to measure the wire resistance of cables, the contact resistance of switches, connectors, relays, the resistance of coils, motors, transformer windings, and metal riveting resistance, the connection resistance test between metal components, the low-value resistance test, and the ground grid and ground poles. Resistance test of connecting conductor, contact resistance test.
The DC low resistance meter is composed of the host, monitoring software, test line, communication line, etc. The large-screen LCD display of the host is clear at a glance. It can store 500 sets of data, and the resistance measurement range: 0.001mO~300.0KO. The upper computer software has functions such as historical data reading, consulting, saving, and reporting.
Model Number Measurement range Accuracy Resolution
6310B 0.001mO~10.000mO Within 18 ? ~28 ?, below 70%rh: ± 0.1%FS ± 20dgt 0.001mO
10.01mO~100.00mO 0.01mO
0.1001O~1.0000O 0.0001O
1.001O~10.000O 0.001O
10.01O~100.00O 0.01O
0.1001KO~1.0000KO 0.0001KO
1.001KO~10.000KO 0.001KO
10.01KO~100.00KO 0.01KO
100.1KO~300.0KO 0.1KO
6310A 0.01mO~100.00mO Within 18 ? ~28 ?, below 70%rh: ± 0.1%FS ± 20dgt 0.01mO
0.1001O~1.0000O 0.0001O
1.001O~10.000O 0.001O
10.01O~100.00O 0.01O
0.1001KO~1.0000KO 0.0001KO
1.001KO~10.000KO 0.001KO
10.01KO~30.00KO 0.01KO
Note: Within 18 ? ~28 ?, above 70%rh, -10 ? ~ within 50 ?, below 80%rh: ± 0.2%FS ± 20dgt.
Model difference
Model Number 6310A 6310B
Resor range 0.01mO~30.00KO 0.001mO~300.0KO
Resolution 10uO 1uO
Data Storage Group 500 Group 500
USB data upload function None Yes
Built-in large-capacity charging battery Yes: Yes:
Real-time display of the battery power Yes: Yes:
Backlight Yes: Yes:
Low Power prompt Yes: Yes:
Overpressure protection Yes: Yes:
Technical specifications
Function It is mainly used to measure the wire resistance of cable, switch, connector, relay contact resistance, coil, motor, transformer winding resistance, metal riveting resistance, connection resistance test between metal components, low value resistance test, resistance test of connecting conductor between ground poles, contact resistance test, etc.
Detection method Four-line method test
Test current current = 1A
Open-circuit voltage = 4.2V
Power: Measured power: = 8W
Power supply DC 3.7V 2000mAh High Capacity lithium battery
Backlight Controlled grey and white screen backlight for dim places
Display mode LCD display, gray and white screen backlight
LCD Size Length and width: 71 mm × 52mm
Instrument size Length and height: 187 mm × 191 mm × 51mm
Test line length About 70cm, Red, Black, 1 bar each
Measurement time About 2 times / second
USB interface With a micro USB interface
Communication line 1 micro USB communication line
Data Storage 500 Group, "MEM" storage indication, display "FULL" symbol means the storage is full
Data access Data access function: "MR" symbol display
Overflow display Overrange overflow function: "OL" Symbol Display
Battery voltage Real-time display of battery power, prompt timely charging when the battery voltage is low
Automatic shutdown "APO" indication, automatic shutdown, 15 minutes after startup
Power consumption Standby: about 100mA (backlight off)
Power-on and open backlight: about 105mA
Measurement: 2A Max
Quality Instrument: 480g(with battery)
Test line: 250g
Working temperature and humidity 18 ? ~28 ?; under 70%rh
Storage temperature and humidity -15 ? ~50 ?; below 70%rh
Overload protection AC 220V/0.0001 seconds between C1-C2 P1-P2 ports, after overload protection, need to restart the instrument to conduct normal test
Insulation resistance Above 10MO (500V between circuit and housing)
Pressure resistance Between the AC 3700V/rms(circuit and the housing)
Electromagnetic characteristics IEC61010-4-3, wireless frequency electromagnetic field = 1V/m
Suitable for safety regulations IEC61010-1, CAT ? 600V, Code for Pollution Level 2, JJG724-1991« DC Digital Ohms, Code for JJG166-1993« DC resistor, and Code for DL/T967-2005 Circuit Resistance Tester and DC Resistance Quick Tester